Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/15/2015 | 1615 | 09/15/2015 | 1658 | Incorrect |
09/16/2015 | 1900 | 09/18/2015 | 1508 | Incorrect |
Subject: | MAO/CPC/S1 - Inlet disconnected |
DataStreams: | maoaoscpcuS1.a1 |
Description: | The instrument inlet was disconnected from ambient sample lines, so the instrument was sampling cabin air. The CPC sheath air line was also disconnected while the capillary flow rate was being measured. In addition, the instrument was taken out of service to examine the capillary more closely, drain butanol, change saturator wick and other maintenance. The capillary was cleaned and during that process the capillary came loose and had to be glued back in with silicone adhesive. |
Suggestions: | Any data available during these time periods should not be used. |
Measurements: | maoaoscpcuS1.a1: more
|