Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/22/2015 | 0000 | 08/04/2015 | 2359 | Incorrect |
Subject: | SGP/AERI/C1 - Reprocess: Spectral calibration to change |
DataStreams: | sgpaerich2C1.a2, sgpaeri01ch2C1.a1, sgpaeri01C1.00, sgpaeriC1.00, sgpaerixC1.00, sgpaerisummaryC1.a1, sgpaeriengineerC1.b1, sgpaeri01engineerC1.a1, sgpaerisummaryC1.a2, sgpaeri01summaryC1.a1, sgpaerich1C1.b1, sgpaerisummaryC1.b1, sgpaeri01ch1C1.a1, sgpaerich1C1.a1, sgpaerich2C1.b1, sgpaerich2C1.a1, sgpaerich1C1.a2 |
Description: | The AERI interferometer laser was changed in the field causing the spectral calibration to change. Effects will be more noticeable for higher wavenumber data. Unless doing trace gas retrievals likely won't notice much of a change once the reprocessed data is in place. |
Suggestions: | Use data with caution. |
Measurements: | sgpaerixC1.00:
sgpaerich1C1.b1:
sgpaerich1C1.a1:
sgpaerich1C1.a2:
sgpaeri01ch2C1.a1:
sgpaeri01engineerC1.a1:
sgpaerisummaryC1.a1:
sgpaeriC1.00:
sgpaerisummaryC1.b1:
sgpaerisummaryC1.a2:
sgpaeri01ch1C1.a1:
sgpaerich2C1.b1:
sgpaeri01summaryC1.a1:
sgpaeriengineerC1.b1:
sgpaerich2C1.a2:
sgpaerich2C1.a1:
sgpaeri01C1.00:
|