DQRID : D151030.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/03/2015110010/15/20151335Suspect
more
Subject:
MAO/CPC/S1 - Environmental, Instrument
DataStreams:maoaoscpcuS1.a1
Description:
A suspected obstruction in the capillary of the cpcu (TSI 3776) caused an error in the 
capillary pressure drop measurement.  This type of problem causes the instrument to report 
flow rates (and therefore particle number concentrations) higher than are measured with a 
flow meter.  This reduced flow rate and the associated error in counting causes the 
instrument to report values lower than that actual atmospheric particle number concentration.
Suggestions: 
Compare against cpcf and total SMPS data for this time period.
Measurements:maoaoscpcuS1.a1:
  • concentration
more

Close this window