Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2014 | 1351 | 04/26/2015 | 1016 | Suspect |
Subject: | ENA/AOS/C1 - Impactor size switch not connected |
DataStreams: | enaaosimpactorC1.a1 |
Description: | Impactor installed and plumbed, HOWEVER, the impactor's 'size read' switch NOT connected. |
Suggestions: | The correct state can be determined by reviewing the Nephelometer pressure (10um=higher pressure, 1um=lower pressure) |
Measurements: | enaaosimpactorC1.a1: more
|