Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/15/2015 | 1600 | 12/31/2015 | 1200 | Incorrect |
Subject: | OLI/RL/M1 - Bad and missing aerosol data |
DataStreams: | olirlM1.a0 |
Description: | During this period, the "aerosol low" and "aerosol high" data channels are bad (PMT high-voltage supply died); the "depol" channel continued to work properly. |
Suggestions: | Do not use data |
Measurements: | olirlM1.a0: more
|