Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
09/08/2016 | 0024 | 11/24/2016 | 1730 | Suspect |
Subject: | ENA/NEPHELOMETER/C1 - Inlet flow changed |
DataStreams: | enaaosnephdry1mC1.b1, enaaosnephdryC1.a1, enaaosnephwet1mC1.b1, enaaosnephwetC1.a1, enaaosnephwetC1.b1, enaaosnephdryC1.b1, enaaoshumidgrC1.01, enaaoshumidgrC1.00 |
Description: | Due to a failure of another instrument on the same sample line, the sample flow for the nephelometers/humidigraph was different than specified. This potentially affected the humidigraph RH and the impactor cut point. Full effect on data quality is unknown, but is likely limited to the impactor cut point being slightly different than expected. |
Suggestions: | Use data with caution |
Measurements: | enaaoshumidgrC1.01: more
enaaosnephdryC1.a1: more enaaosnephwetC1.a1: more enaaosnephdryC1.b1: more enaaosnephdry1mC1.b1: more enaaosnephwetC1.b1: more enaaosnephwet1mC1.b1: more enaaoshumidgrC1.00: more |