DQRID : D161230.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
10/24/2016210410/27/20162345Incorrect
more
Subject:
NSA/NEPHELOMETER/X1 - Signal noise from NEPH PCB electrical problem
DataStreams:nsaaosnephdry1mX1.b1, nsaaosnephdryX1.a1, nsaaosnephdryX1.b1
Description:
The instrument scattering signal had large spikes just prior to removal for repair.
Suggestions: 
Remove spikes from data before using
Measurements:nsaaosnephdryX1.b1:
  • Bbs_R_Dry_Neph3W_1
  • RH_Neph_Dry
  • Bbs_B_Dry_Neph3W_1
  • Bs_R_Dry_Neph3W_1
  • P_Neph_Dry
  • Bs_B_Dry_Neph3W_1
  • Bbs_R_Dry_Neph3W
  • Bs_G_Dry_Neph3W_1
  • dilution_correction_factor
  • Bbs_B_Dry_Neph3W
  • seconds_after_transition
  • Bbs_G_Dry_Neph3W_1
  • RH_Neph_Inlet_Dry
  • T_Neph_Dry
  • Bs_G_Dry_Neph3W
  • density_correction_factor
  • Bs_R_Dry_Neph3W
  • Bbs_G_Dry_Neph3W
  • T_Neph_Inlet_Dry
  • Bs_B_Dry_Neph3W
more
nsaaosnephdryX1.a1:
  • Bbs_R_Dry_Neph3W
  • Bbs_R_Dry_10um_Neph3W_1
  • Bbs_R_Dry_1um_Neph3W_1
  • RH_Neph_Inlet_Dry
  • T_NephVol_Dry
  • Bbs_G_Dry_10um_Neph3W_1
  • T_Neph_Inlet_Dry
  • Bs_G_Dry_1um_Neph3W_1
  • RH_Neph_Vol_Dry
  • Bs_R_Dry_Neph3W
  • Bbs_G_Dry_Neph3W
  • Bs_G_Dry_Neph3W
  • P_Neph_Dry
  • Bs_G_Dry_10um_Neph3W_1
  • RH_NephInlet_Dry
  • Bbs_B_Dry_Neph3W
  • f2_flags
  • Bs_R_Dry_1um_Neph3W_1
  • T_Neph_Vol_Dry
  • Bs_B_Dry_10um_Neph3W_1
  • Bs_B_Dry_1um_Neph3W_1
  • Bs_B_Dry_Neph3W
  • RH_NephVol_Dry
  • Bbs_B_Dry_1um_Neph3W_1
  • T_NephInlet_Dry
  • Bs_R_Dry_10um_Neph3W_1
  • Bbs_G_Dry_1um_Neph3W_1
  • Bbs_B_Dry_10um_Neph3W_1
more
nsaaosnephdry1mX1.b1:
  • Bbs_B_Dry_Neph3W_raw
  • Bs_B_Dry_Neph3W
  • T_Neph_Dry
  • P_Neph_Dry
  • Bbs_G_Dry_Neph3W
  • RH_Neph_Dry
  • Bbs_G_Dry_Neph3W_raw
  • Bs_R_Dry_Neph3W_raw
  • T_Neph_Inlet_Dry
  • Bbs_B_Dry_Neph3W
  • Bbs_R_Dry_Neph3W_raw
  • Bs_B_Dry_Neph3W_raw
  • Bs_R_Dry_Neph3W
  • Bs_G_Dry_Neph3W_raw
  • Bs_G_Dry_Neph3W
  • Bbs_R_Dry_Neph3W
more

Close this window