ENA/ACSM/C1 - Instrument malfunction
DataStreams:enaaosacsmC1.01, enaaosacsmC1.00, enaaosacsmC1.b1, enaaosacsmC1.a1
Description:
The instrument interlock was activated shutting the inlet valve and turning off voltages 
to the filament and detector. The data acquisition was unaffected and data were collected, 
however there was no signal. Instrument was restarted and is now collecting data.
Suggestions: 
Do not use data
Measurements:enaaosacsmC1.b1:
  • total_organics
  • chloride
  • org_mx
  • amus
  • ammonium
  • nitrate