DQRID : D170508.7
Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/23/2017000004/19/20171900Missing
more
Subject:
SGP/MPLCMASK/C1 - Reprocess: Config file change caused missing data
DataStreams:sgp30smplcmask1zwangC1.c1
Description:
The manufacturer changed the laser fire bin setting in the instrument config file after 
maintenance on the instrument. The data is fine, but the VAP was not setup to handle this 
change and did not run during this time period. The VAP has now been updated to handle 
this configuration more flexibly and can now be rerun to provide the missing data.
Suggestions: 
Wait until reprocessed data becomes available
Measurements:sgp30smplcmask1zwangC1.c1:
  • cloud_base
  • linear_depol_ratio
  • lon
  • qc_cloud_mask
  • cloud_top_attenuation_flag
  • qc_linear_depol_ratio
  • deadtime_correction
  • background_signal
  • cloud_top_layer
  • cloud_base_layer
  • qc_backscatter
  • cloud_top
  • backscatter
  • qc_backscatter_snr
  • afterpulse_correction
  • num_cloud_layers
  • base_time
  • cloud_mask
  • time_offset
  • linear_depol_snr
  • overlap_correction
  • height
  • qc_linear_depol_snr
  • deadtime_correction_counts
  • backscatter_snr
  • lat
  • shots_summed
  • time
  • alt
more

Close this window