DQRID : D170511.12
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/01/2017070905/01/20171329Missing
more
Subject:
ENA/ACSM/C1 - Filament burned out
DataStreams:enaaosacsmC1.01, enaaosacsmC1.00, enaaosacsmC1.b1, enaaosacsmC1.a1
Description:
Filament #1 burned out 2017-05-11 7:09. Filament #2 was turned on at 2017-05-11 13:29. 
Mass spectra were collected, but there were no ions generated so there was no signal. Only 
detector noise was collected.
Suggestions: 
Measurements:enaaosacsmC1.b1:
  • qc_org_mx_err
  • chloride
  • qc_chloride
  • time_bounds
  • lon
  • amus
  • time_offset
  • ammonium
  • time
  • org_mx_err
  • qc_ammonium
  • qc_sulfate
  • qc_nitrate
  • total_organics
  • alt
  • org_mx
  • qc_org_mx
  • qc_total_organics
  • nitrate
  • lat
  • base_time
  • sulfate
more
enaaosacsmC1.a1:
  • time_bounds
  • alt
  • ammonium
  • total_organics
  • inlet_closed
  • sulfate
  • nitrate
  • lat
  • vaporizer
  • flow_rate
  • lon
  • base_time
  • emission_current
  • time_offset
  • time
  • chloride
  • inlet_open
more
enaaosacsmC1.00:
  • Raw data stream - documentation not supported
more
enaaosacsmC1.01:
  • Raw data stream - documentation not supported
more

Close this window