Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/05/2000 | 1800 | 01/25/2001 | 2016 | Suspect |
Subject: | SGP/MWR/B4 - noisy data |
DataStreams: | sgpmwrlosB4.a1, sgpmwrtipB4.a1, sgpmwrlosB4.b1 |
Description: | The MWR exhibited occasional negative sky brightness temperatures that were resolved, after eliminating the instrument and communication lines as the source of the problem, by reformatting and reconfiguring the laptop computer. |
Suggestions: | |
Measurements: | sgpmwrlosB4.a1: more
sgpmwrtipB4.a1: more sgpmwrlosB4.b1: more |