Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/03/2001 | 2328 | 12/04/2001 | 0158 | Incorrect |
Subject: | SGP/SONDE/C1 - Bad temp (other?) data in 12/3/01:2328 sounding |
DataStreams: | sgpsondewnpnC1.00, sgpsondeC1.00, sgpsondewnpnC1.b1, sgpsondewnpnC1.a1 |
Description: | It looks as though the temperature sensor was damaged on launch. The recorded temp goes from 16.4 at the surface to -42.6 at 2 seconds into the flight. The temperature never exceeds -42 though the RH and pressure look reasonable, but probably should be considered questionable. |
Suggestions: | |
Measurements: | sgpsondeC1.00: more
sgpsondewnpnC1.b1: more sgpsondewnpnC1.a1: more sgpsondewnpnC1.00: more |