DQRID : D011211.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/10/2000000006/30/20002359Incorrect
more
Subject:
SGP/SIRS/C1 - Reprocess: Wire are switched
DataStreams:sgpsirsC1.a1
Description:
Upwelling longwave and downwelling longwave values are switched.  Instrument Engineering 
logs (OMIS) indicate the following:  "T/S indicated that the readings may be caused by the 
sensor wires being switched around.  Switched the wires labeled 5H with the wire labeled 
5L on the data logger.  The readings on the DIR went from minus readings to positive 
readings.  Will await further instruction from site scientist to perform any other PM".
Suggestions: 
Reprocess the data.
Measurements:sgpsirsC1.a1:
  • up_long_hemisp
  • down_long_hemisp
more

Close this window