DQRID : D020129.13
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/06/2002193001/06/20022101Suspect
01/09/2002033001/18/20020001Suspect
01/08/2002173001/08/20021801Suspect
01/05/2002193001/05/20021931Suspect
10/12/2001173510/18/20011601Suspect
11/22/2001013011/25/20011531Suspect
01/07/2002233001/07/20022331Suspect
more
Subject:
SGP EBBR E13 - E13 EBBR Left T/RH Probe Temperature Incorrect
DataStreams:sgp30ebbrE13.b1, sgp15ebbrE13.b1
Description:
The left T/RH Probe Temperature was often incorrect, causing the other indicated
measurements to be incorrect also.  

When the home30 signal had an offset, the top RH probe temperature was incorrect
but not the bottom RH probe temperature.
Suggestions: 
The incorrect measurements cannot easily be recalculated.
Measurements:sgp15ebbrE13.b1:
  • temp_trh_left_rr
more
sgp30ebbrE13.b1:
  • bowen_ratio
  • temp_trh_top
  • sensible_heat_flux
  • latent_heat_flux
more

Close this window