DQRID : D030117.16
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/04/2002171509/04/20021730Incorrect
10/13/2002130010/15/20021930Incorrect
10/15/2002214510/16/20021630Incorrect
10/13/2002123010/13/20021234Incorrect
more
Subject:
SGP/EBBR/E22 - AEM Hung between home positions
DataStreams:sgp30ebbrE22.b1, sgp5ebbrE22.b1, sgp15ebbrE22.b1
Description:
When the AEM (Automatic Exchange Mechanism) is hung 
inbetween home positions, the gradients of temperature and 
relative humidity, and thus sensible and latent heat fluxes, 
are incorrect.

There is no work-around to this problem.
Suggestions: 
Do not use data
Measurements:sgp15ebbrE22.b1:
  • temp_air_right
  • temp_air_left
more
sgp5ebbrE22.b1:
  • temp_trh_bottom
  • rh_bottom_fraction
  • vapor_pressure_top
  • temp_air_top
  • temp_air_bottom
  • rh_top_fraction
  • vapor_pressure_bottom
  • temp_trh_top
more
sgp30ebbrE22.b1:
  • temp_trh_bottom
  • rh_top_fraction
  • temp_air_top
  • rh_bottom_fraction
  • latent_heat_flux
  • vapor_pressure_top
  • temp_trh_top
  • temp_air_bottom
  • bowen_ratio
  • sensible_heat_flux
  • vapor_pressure_bottom
more

Close this window