DQRID : D030117.8
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/09/2002173012/11/20021930Does not affect quality
more
Subject:
SGP/EBBR/E20 - Reprocessed and Corrected: Net Rad Coefficients Incorrect
DataStreams:sgp30ebbrE20.b1, sgp5ebbrE20.b1
Description:
DATA REPROCESSED and CORRECTED: 

Incorrect calibration resulted in bad q (net radiation) values. Simple multipliers 
corrected the data and then sensible and latent heat fluxes were recalculated.

The data were archived in March 2017. Please retrieve the reprocessed data for use in your 
analysis.

You may find these data using the ARM Discovery tool at the following link:
http://www.archive.arm.gov/discovery/#v/results/s/finst::ebbr/fsite::sgp/ffac::sgp.E20 

ORIGINAL DQR CATEGORY: Red( Incorrect)

REPROCESSED and CORRECTED DATA ARCHIVED:2017-02

REPROCESS IDENTIFICATION NUMBER: RID 231

---------------------------------------------------
--- ORIGINAL DQR DESCRIPTION ---
---------------------------------------------------
The Net Radiometer coefficients were incorrect in the CR10
program when the replacement EBBR unit was installed on
08/09/02.

The data can be easily corrected; see the Suggestion area below.
Suggestions: 
Measurements:sgp30ebbrE20.b1:
  • latent_heat_flux
  • net_radiation
  • sensible_heat_flux
more
sgp5ebbrE20.b1:
  • net_radiation
more

Close this window