DQRID : D030909.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/11/2003000008/28/20032359Suspect
more
Subject:
SGP/MWR/E14  - thermal instability
DataStreams:sgpmwrlosE14.b1, sgpmwrlosE14.a1
Description:
During periods of high ambient temperature, the instrument may have been thermally 
unstable.
Suggestions: 
The data are suspect when the blackbody temperature, tkbb, exceeds 318 K and the air 

temperature, tkair, exceeds 311 K.
Measurements:sgpmwrlosE14.b1:
  • tbsky23
  • tbsky31
  • liq
  • vap
more
sgpmwrlosE14.a1:
  • vap
  • liq
  • tbsky31
  • tbsky23
more