Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/11/2003 | 0000 | 08/28/2003 | 2359 | Suspect |
Subject: | SGP/MWR/E14 - thermal instability |
DataStreams: | sgpmwrlosE14.b1, sgpmwrlosE14.a1 |
Description: | During periods of high ambient temperature, the instrument may have been thermally unstable. |
Suggestions: | The data are suspect when the blackbody temperature, tkbb, exceeds 318 K and the air temperature, tkair, exceeds 311 K. |
Measurements: | sgpmwrlosE14.b1: more
sgpmwrlosE14.a1: more |