DQRID : D050811.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
08/02/2005112908/02/20051225Incorrect
more
Subject:
PYE/SONDE/M1 - Broken RH sensor
DataStreams:pyesondewnpnM1.b1
Description:
One of the RH sensors on this radiosonde was bad.  The RH oscillates rapidly between near 
0 and ambient up to approximately 11 km.  The near-zero values are incorrect.
Suggestions: 
Low-pass filter the RH data for the first hour of the sounding.
Measurements:pyesondewnpnM1.b1:
  • rh
  • dp
more