DQRID : D050920.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/13/2005000010/18/20052000Does not affect quality
more
Subject:
NSA/MWR/C1 - Instrument problem, Heater problem
DataStreams:nsamwrlosC1.b1, nsamwrC1.00
Description:
The heater of the MWR was ON all the time even when the moisture sensor was dry. Jim 
Liljegren suspected a problem with the relay that activates the heater. The suggested action 
was to send the faulty assembly to SGP for repair and get a spare unit from SGP.
The spare unit was sent on 9/15 and was installed on 10/18.
Suggestions: 
Data should not be affected by this problem
Measurements:nsamwrC1.00:
  • Raw data stream - documentation not supported
more
nsamwrlosC1.b1:
  • liq
  • actaz
  • time_offset
  • bb31
  • tkxc
  • tknd
  • lat
  • tnd_nom23
  • tbsky31
  • actel
  • tnd23
  • vap
  • sky31
  • alt
  • bbn23
  • tkair
  • sky23
  • sky_ir_temp
  • tc31
  • tbsky23
  • tnd_nom31
  • wet_window
  • tc23
  • lon
  • ir_temp
  • tkbb
  • base_time
  • bb23
  • bbn31
  • tnd31
more