DQRID : D051106.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/14/2005000009/20/20052000Incorrect
more
Subject:
SGP/EBBR/E7 - Net Radiation Incorrect; Sensible and Latent Heat Fluxes Incorrect
DataStreams:sgp15ebbrE7.b1, sgp5ebbrE7.b1, sgp30ebbrE7.b1
Description:
The top dome of the net radiometer was cracked; net radiation, and sensible and latent 
heat flux measurements were incorrect.
Suggestions: 
Measurements:sgp5ebbrE7.b1:
  • net_radiation
more
sgp15ebbrE7.b1:
  • net_radiation_mv
more
sgp30ebbrE7.b1:
  • latent_heat_flux
  • net_radiation
  • sensible_heat_flux
more