TWP/SKYRAD/C1  - IRT failure
DataStreams:twpskyrad60sC1.b1
Description:
A moisture related fault was found in the IRT cable after a day of very heavy rain. The 
cable was replaced but the internal power supply was apparently damaged. The instrument was 
replaced with a spare IRT.
Suggestions: 
Measurements:twpskyrad60sC1.b1:
  • sky_ir_temp_min
  • sky_ir_temp_std
  • sky_ir_temp
  • sky_ir_temp_max