Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/21/2005 | 1500 | 12/05/2005 | 0100 | Incorrect |
Subject: | TWP/GNDRAD/C3 - IRT malfunction |
DataStreams: | twpgndrad60sC3.b1 |
Description: | The internal IRT power supply failed (probably due to a lightning strike on 11/14) and was drawing excessive current on the data logger's DC system. The problem was corrected by replacing the instrument with a spare. |
Suggestions: | Do not use IRT data. |
Measurements: | twpgndrad60sC3.b1: more
|