Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
11/16/2005 | 2000 | 02/07/2006 | 2030 | Incorrect |
Subject: | SGP/MFRSR/E1 - Instrument not shaded properly |
DataStreams: | sgpmfrsrE1.a0, sgpmfrsrlangleyE1.c1, sgpmfrsrE1.b1, sgpmfrsrlangplotE1.c1 |
Description: | Shading issues throughout this period. It is first noticeable on 11/16/2005, albeit ever so slightly. Maintenance was attempted on 1/10/2006 but that only shifted the shading problem from the afternoon to a mostly all day situation. And it became much worse. Maintenance on 2/7/2006 finally fixes the alignment. If data from this period is critical, the morning data from 11/16/2005 to 1/09/2006 is probably OK. Don't use any diffuse or direct data from 1/10/2006 through 2/7/2006 at 2030. |
Suggestions: | |
Measurements: | sgpmfrsrlangleyE1.c1: more