DQRID : D060420.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/15/2006000006/03/20060000Suspect
more
Subject:
NIM/MWRP/M1 - IRT Sensor Calibration
DataStreams:nimmwrpM1.00, nimmwrprawM1.00, nimmwrpM1.b1
Description:
Discrepancies between the MWRP and Skyrad IRT were observed in January. The MWRP IRT 
readings were constantly higher than those of the skyrad IRT. Some changes were introduced to 
the MWRP IRT to address the problem (see DQR D060602.2). On May 3, the IRT mirror was 
replaced. After the change in the mirror, the MWRP IRT readings became about 5-8 C lower. On 
June 3 the agreement between MWRP and Skyrad IRT became satisfactory (2-5 degree C 
difference).

It is hard to numerically quantify the difference in the IRT readings caused by the change 
in the mirror. We can only warn the user that between January 1 and June 3 the MWRP 
readings are 8 to 15 degree higher than the skyrad IRT.
Suggestions: 
use data with caution