Start DateStart TimeEnd DateEnd Time Data Quality Metric
03/25/2006003704/11/20062120Incorrect
more
Subject:
NSA/MET/C1 - CMH Data Incorrect
DataStreams:nsametC1.b1
Description:
CMH data was bad due to incorrect calibration caused by procedural errors. The procedures 
were corrected and a new calibration completed.
Suggestions: 
Do not use data. Use 2m Temp and RH data.
Measurements:nsametC1.b1:
  • cmh_rh
  • cmh_sat_vapor_pressure
  • cmh_vapor_pressure
  • cmh_temp
  • cmh_dew_point