Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
05/11/2006 | 2008 | 09/25/2006 | 2359 | Does not affect quality |
Subject: | SGP/NIMFR/C1 - Reprocessed: Miscalibrated detectors #5 and #6 |
DataStreams: | sgpnimfrC1.b1 |
Description: | Filter detectors 5 and 6 were miscalibrated following repair and reinstallation in May 2006. The data for 20060511-20060928 were reprocessed and archived 20061116. |
Suggestions: | Request the reprocessed data from the archive for use in your analysis. |
Measurements: | sgpnimfrC1.b1: more
|