DQRID : D070709.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/19/2004021406/02/20072359Does not affect quality
more
Subject:
NSA/MWRP/C1 - Reprocessed: 6-channel retrieval replaced by 5-channel retrieval
DataStreams:nsamwrpC1.b1
Description:
Due to repeated failure of the V-band receiver and calibration problems, the 6-channel 
retrieval was retired and a new 5-channel retrieval which uses only the K-band channels was 
implemented.

All data previously processed with the 6-channel retrieval have been reprocessed.

Please note that the 7 channels (brightness temperatures) in the V-band range (51 to 60 
GHz) should not be used.  See D080331.1 and D080331.2.
Suggestions: 
Retrieve the reprocessed data for use in your analysis
Measurements:nsamwrpC1.b1:
  • brightnessTemperature
  • liquidWaterPath
  • liquidWaterPath2
more