Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
12/17/2007 | 0000 | 02/13/2008 | 2330 | Does not affect quality |
Subject: | NSA/AERI/C1 - Reprocessed: Failed temp sensor affected calibration |
DataStreams: | nsaaerich1C1.b1, nsaaerich2C1.b1, nsaaerisummaryC1.b1 |
Description: | The blackbody support structure temperature sensor on the NSA.C1 AERI were biased high by approximately 25K beginning 20071217 and became progressively worse until the sensor was replaced 20080213. The temperature from this sensor is used to calibrate the radiance data for the AERI. The AERI data have been reprocessed by replacing the data from the failed sensor with the cold blackbody temperature sensor data and producing corrected/recalibrated radiance data. The reprocessed data were archived in March 2008. |
Suggestions: | Retrieve reprocessed data set from ARM Archive. |
Measurements: | nsaaerich1C1.b1: more
nsaaerisummaryC1.b1: more nsaaerich2C1.b1: more |