DQRID : D100128.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/23/2009160001/27/20101510Incorrect
more
Subject:
SGP/SIRS/E7 - Upwelling Longwave inconsistent
DataStreams:sgpsirsE7.b1, sgpsirs20sE7.a0
Description:
The upwelling IR PIR instrument (up_long_hemisp) had bad connections during this time 
period.  The case thermistor (Mux set #2) was reading erratic values and corrupting the 
up_long_hemisp data completely. 

The net_ir was valid but cannot be corrected without the case thermistor measurement to 
determine the incoming IR (i.e. up_long_hemisp).
Suggestions: 
Measurements:sgpsirs20sE7.a0:
  • inst_up_long_dome_resist
  • inst_up_long_hemisp_tp
  • inst_up_long_case_resist
more
sgpsirsE7.b1:
  • inst_up_long_case_temp
  • inst_up_long_dome_temp
  • up_long_netir
  • up_long_hemisp
  • up_long_hemisp_min
  • up_long_hemisp_std
  • up_long_hemisp_max
more

Close this window