DQRID : D110511.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
04/04/2011130004/27/20111100Incorrect
more
Subject:
NSA/MWRP/C1 - Incorrect brightness temperatures
DataStreams:nsamwrpC1.b1
Description:
The radiometer was reinstalled on 4/3/2011 but the 30 GHz data immediately appeared 
extremely noisy. The problem appears related to noise diode instability.
Suggestions: 
Do not use data and retrievals during the indicated time
Measurements:nsamwrpC1.b1:
  • liftingCondensationLevel
  • waterVaporMixingRatio
  • brightnessTemperature
  • dewpointTemperature
  • equilibriumLevel
  • totalPrecipitableWater2
  • temperature
  • relativeHumidity
  • liquidWaterContent
  • virtualTemperature
  • totalPrecipitableWater
  • levelFreeConvection
  • waterVaporDensity
  • equilibriumLevelPres
  • levelFreeConvectionPres
  • liftingCondensationLevelPres
  • liquidWaterPath
  • liquidWaterPath2
more