DQRID : D111010.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
09/02/2011002209/24/20110145Suspect
more
Subject:
NSA/SKYRAD/C1 - Noisy data due to grounding issue
DataStreams:nsaskyrad60sC1.b1, nsaqcrad1longC1.c1
Description:
Improper grounding of the IRT serial cable connection caused significant noise in the 
signals of the other SKYRAD instruments. It was corrected by disconnecting the serial cable 
from the RocketPort.
Suggestions: 
Measurements:nsaskyrad60sC1.b1:
  • down_short_diffuse_hemisp
  • down_long_hemisp2
  • inst_down_long_hemisp2_tp
  • down_short_hemisp_max
  • down_short_hemisp_std
  • down_short_hemisp
  • down_short_diffuse_hemisp_min
  • inst_down_long_hemisp2_dome_temp
  • down_long_hemisp1_std
  • short_direct_normal_max
  • down_long_hemisp1_max
  • short_direct_normal_std
  • inst_down_long_hemisp1_case_temp
  • down_long_hemisp2_max
  • down_long_hemisp2_std
  • down_short_diffuse_hemisp_std
  • short_direct_normal
  • down_short_diffuse_hemisp_max
  • down_long_hemisp1_min
  • short_direct_normal_min
  • down_long_hemisp2_min
  • inst_down_long_hemisp1_tp
  • inst_down_long_hemisp2_case_temp
  • inst_down_long_hemisp1_dome_temp
  • down_long_hemisp1
  • down_short_hemisp_min
more
nsaqcrad1longC1.c1:
  • down_short_diffuse_hemisp
  • down_long_hemisp
  • down_short_hemisp
  • short_direct_normal
more

Close this window