DQRID : D111212.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
11/08/2011202512/12/20112014Does not affect quality
more
Subject:
SGP/IRT/E38 - Reprocessing Notice (RID 726)
DataStreams:sgpirt200msE38.a1, sgpirtE38.b1
Description:
See below for details of reprocessing that was completed on 20140822.  Original DQR 
category: 4 (Red, Incorrect ).

-----------------------------------------------------------

The IRT was measuring in units of C instead of K because the IRT was not configured 
correctly by the software when started. The units parameter was changed manually.

Data will be reprocessed
Suggestions: 
Measurements:sgpirtE38.b1:
  • sky_ir_temp
  • ref_ir_temp
  • sky_ir_temp_min
  • sky_ir_temp_max
more
sgpirt200msE38.a1:
  • sky_ir_temp
  • ref_ir_temp
more

Close this window