Description: | See below for details of reprocessing that was completed on 20150213. Original
DQR category: 4 (Red, Incorrect).
-----------------------------------------------------------
Applied calibration coefficient and serial number for down_short_diffuse was incorrect for
the specified period.
Perform the following calculation to correct the data,
DD= downwelling diffuse instrument:
Old Calibration coefficient = 114.74[W/m2\mV]
New Calibration Coefficient = 109.51[W/m2\mV]
DD(new)[W/m2]=( DD(old)[W/m2]/ 114.74 [W/m2\mV] )* 109.51 [W/m2/mV] |