DQRID : D120124.5
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/01/2011000012/10/20110000Incorrect
more
Subject:
TWP/MWR3C/C3 - Incorrect calibration
DataStreams:twpmwr3cC3.b1
Description:
When the radiometer was relocated to the TWP site the 89-GHz detector non-linearity 
coefficient got somehow erased from the radiometer memory. The radiometer was therefore unable 
to calibrate properly.
Suggestions: 
Measurements:twpmwr3cC3.b1:
  • tbsky23
  • tbsky89
  • lwp
  • pwv
  • tbsky30
more