DQRID : D120414.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/12/2006110007/21/20061635Incorrect
more
Subject:
SGP/QCRAD/E9 - Tracker not working
DataStreams:sgpqcrad1longE9.c2, sgpqcrad1longE9.s1, sgpqcrad1longE9.c1, sgpqcrad1longE9.s2
Description:
Tracker appears to be out of alignment or inoperative during these times. The 
short_direct_normal (NIP) and down_short_diffuse_hemisp (8-48) data apear to be bad and not 
recoverable.

Documented in instrument DQRs: D060718.2 & D070312.1
Suggestions: 
Don't use SW direct or diffuse data during these time periods. If 

"source_BestEstimate_down_short_hemisp" is 0, don't use Best Estimate GSW data. Note that fewer quality 
checks can be applied accurately to the BestEstimate data.
Measurements:sgpqcrad1longE9.c1:
  • BestEstimate_down_short_hemisp
  • short_direct_normal
  • down_short_diffuse_hemisp
more
sgpqcrad1longE9.s1:
  • BestEstimate_down_short_hemisp
  • short_direct_normal
  • down_short_diffuse_hemisp
more
sgpqcrad1longE9.c2:
  • BestEstimate_down_short_hemisp
  • short_direct_normal
  • down_short_diffuse_hemisp
more
sgpqcrad1longE9.s2:
  • short_direct_normal
  • BestEstimate_down_short_hemisp
  • down_short_diffuse_hemisp
more

Close this window