DQRID : D120510.23
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/02/2001220006/12/20011600Suspect
10/26/1997230001/06/19981700Suspect
06/30/1997030009/08/19971800Suspect
more
Subject:
SGP/QCRAD/E12  - Tc/Td missing or noisy
DataStreams:sgpqcrad1longE12.s2, sgpqcrad1longE12.c2, sgpqcrad1longE12.c1, sgpqcrad1longE12.s1
Description:
Case and dome temperatures for Longwave fluxes are noisy during this period corresponding 
to additional noise in the longwave fluxes.

No instrument DQRs exist documenting this problem.
Suggestions: 
Longwave fluxes have higher uncertainties during this time period, use with caution.
Measurements:sgpqcrad1longE12.s2:
  • down_long_hemisp
  • LWupTd
  • LWupTc
  • up_long_hemisp
  • LWdnTd
  • LWdnTc
more
sgpqcrad1longE12.c2:
  • LWupTc
  • LWupTd
  • up_long_hemisp
  • LWdnTd
  • LWdnTc
  • down_long_hemisp
more
sgpqcrad1longE12.s1:
  • LWupTc
  • LWupTd
  • LWdnTc
  • LWdnTd
  • down_long_hemisp
  • up_long_hemisp
more
sgpqcrad1longE12.c1:
  • down_long_hemisp
  • up_long_hemisp
  • LWupTc
  • LWupTd
  • LWdnTd
  • LWdnTc
more

Close this window