Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
07/17/1997 | 0000 | 07/30/1997 | 1900 | Suspect |
04/09/1997 | 1400 | 05/07/1997 | 1900 | Suspect |
Subject: | SGP/QCRAD/E18 - LWup Tc/Td bad |
DataStreams: | sgpqcrad1longE18.c1, sgpqcrad1longE18.s2, sgpqcrad1longE18.s1, sgpqcrad1longE18.c2 |
Description: | Case and dome temperatures for upwelling longwave fluxes are bad during this period, so some checks/corrections cannot be applied to the upwelling longwave fluxes. No instrument DQRs exist documenting this problem. |
Suggestions: | Upwelling longwave fluxes have higher uncertainties during this time period, use with caution. |
Measurements: | sgpqcrad1longE18.c1: more
sgpqcrad1longE18.s1: more sgpqcrad1longE18.c2: more sgpqcrad1longE18.s2: more |