DQRID : D130227.6
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/07/2013210002/13/20131846Incorrect
more
Subject:
PVC/SKYRAD/M1 - Tracker removed for storm
DataStreams:pvcskyrad20sM1.a0, pvcskyrad60sM1.b1
Description:
Site preparation for a pending storm in Cape Cod Area included stowing the tracker for the 
affected period.  During this time the only SKYRAD instruments with accurate 
measurements were down_short_hemisp (global horizontal shortwave pyranometer instrument), 
down_long_hemisp1 and down_long_hemisp2 (pyrgeometer instruments).

The short_direct_normal (pyrheliometer DNI instrument) and down_short_diffuse_hemisp (8-48 
shaded diffuse instrument) data are not recoverable.

Additionally, there may have been undocumented periods of missing data during this time.
Suggestions: 
Measurements:pvcskyrad60sM1.b1:
  • down_short_diffuse_hemisp_std
  • short_direct_normal
  • down_short_diffuse_hemisp_max
  • short_direct_normal_std
  • short_direct_normal_min
  • short_direct_normal_max
  • down_short_diffuse_hemisp
  • down_short_diffuse_hemisp_min
more
pvcskyrad20sM1.a0:
  • inst_diffuse
  • inst_direct_normal
more