DQRID : D130903.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
01/11/2011000004/25/20112359Suspect
more
Subject:
SBS/MWR3C/M1 - Software problem-IRT missing
DataStreams:sbsmwr3cM1.b1
Description:
The IRT data appear intermittent. Upon investigation it was found that data are missing 
when the temperature is below the minimum sensitivity of the IRT sensor. The vendor agreed 
to modify the software to better handle data in the low sensitivity range of the 
instrument. The new software will be installed at the next deployment.
Suggestions: 
Measurements:sbsmwr3cM1.b1:
  • infrared_temperature
more