DQRID : D130911.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
07/01/2013150008/14/20130103Incorrect
more
Subject:
TWP/GNDRAD/C2 - Incorrect or missing data
DataStreams:twpgndrad60sC2.b1
Description:
During the time range specified the GNDRAD datalogger and all instruments were reporting 
either missing or incorrect measurement values.  The problem resolved itself on 8/14/13 
with the shortwave, longwave and IRT instruments returning to normal readings.  

Note that on late GMT on 8/18/13 through early 8/19/13 site ops replaced the logger 
backplane resulting in several hours of missing data.  The shortwave (PSP instrument) and 
longwave(PIR instrument) upwelling data were good before and after this time.
Suggestions: 
Data are unrecoverable.
Measurements:twpgndrad60sC2.b1:
  • up_short_hemisp_max
  • inst_up_long_hemisp_dome_temp
  • lon
  • time_offset
  • alt
  • sfc_ir_temp_min
  • base_time
  • up_short_hemisp_min
  • up_long_hemisp
  • vBatt
  • logger_temp
  • sfc_ir_temp
  • up_long_hemisp_min
  • inst_up_long_hemisp_tp
  • lat
  • up_short_hemisp
  • sfc_ir_temp_max
  • time
  • up_long_hemisp_max
  • inst_up_long_hemisp_case_temp
  • sfc_ir_temp_std
  • up_short_hemisp_std
  • up_long_hemisp_std
more