DQRID : D140604.1
Start DateStart TimeEnd DateEnd Time Data Quality Metric
05/19/2014214205/30/20141557Missing
more
Subject:
SGP/SIRS/E15 - Data missing
DataStreams:sgpsirsE15.b1, sgpsirs20sE15.a0
Description:
The Short Direct Normal (NIP) instrument had a broken wire in the connector causing data 
to be missing or incorrect.
Suggestions: 
The data can be approximated from the two other shortwave datastreams available using 

the following relationship:
short_direct_normal = (down_short_hemisp - down_short_diffuse)/cos(zenith angle)
Measurements:sgpsirs20sE15.a0:
  • inst_direct_normal
more
sgpsirsE15.b1:
  • short_direct_normal
  • short_direct_normal_max
  • short_direct_normal_min
  • short_direct_normal_std
more