DQRID : D140625.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/12/2014054406/22/20140612Incorrect
more
Subject:
TMP/AOS/S1 - Impactor state occasionally incorrect
DataStreams:tmpaosimpactorS1.a1
Description:
The computer controlling the impactor state was rebooting every 15 minutes causing 
periodic gaps in the data and interrupting the normal sequence of states.
Suggestions: 
The impactor state can be determined unambiguously by looking at the dry neph pressure.
Measurements:tmpaosimpactorS1.a1:
  • impactor_state