DQRID : D140703.2
Start DateStart TimeEnd DateEnd Time Data Quality Metric
12/30/2013160506/19/20142000Incorrect
more
Subject:
MAO/AOS/S1 - Impactor Setting Reversed
DataStreams:maoaosimpactorS1.a1
Description:
The impactor was reporting the opposite state of what was being measured.  I.e. measuring 
10um but reporting 1um.
Suggestions: 
The correct state can be determined by reviewing the Nephelometer pressure 

(10um=higher pressure, 1um=lower pressure)
Measurements:maoaosimpactorS1.a1:
  • impactor_state
more