Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/11/2014 | 1350 | 05/30/2014 | 0010 | Missing |
Subject: | SGP/CLAP/C1 - Instrument removed to check low signal |
DataStreams: | sgpaosclap3wC1.b1 |
Description: | Instrument was removed to check low signal values. After instrument was reinstalled, redid clap spot size measurement and compared to another CLAP in lab with both on filtered air and both on outside air. Instruments were within 2%. |
Suggestions: | Use PSAP data. |
Measurements: | sgpaosclap3wC1.b1: more
|