DQRID : D140805.3
Start DateStart TimeEnd DateEnd Time Data Quality Metric
06/13/2014215607/30/20142130Incorrect
more
Subject:
OLI/SKYRAD/M1 - Solar tracker not functioning correctly
DataStreams:oliskyrad20sM1.a0, oliskyrad60sM1.b1
Description:
The solar tracker was not functioning correctly during the affected time range. Data from 
instruments is directly reliant on correct tracker operation (short_direct_normal and 
down_short_diffuse) and are not recoverable.
Suggestions: 
Do not use data
Measurements:oliskyrad20sM1.a0:
  • inst_direct_normal
  • inst_diffuse
more
oliskyrad60sM1.b1:
  • short_direct_normal_max
  • short_direct_normal_std
  • short_direct_normal_min
  • short_direct_normal
  • down_short_diffuse_hemisp
  • down_short_diffuse_hemisp_std
  • down_short_diffuse_hemisp_min
  • down_short_diffuse_hemisp_max
more