DQRID : D141023.4
Start DateStart TimeEnd DateEnd Time Data Quality Metric
02/16/2012000010/20/20142359Incorrect
more
Subject:
ENA/MAG/PVC/TMP/CCN/C1 - Reprocess: CCN temperature unstable flag reversed
DataStreams:tmpaosccn100S1.a1, enaaosccn100C1.a1, pvcaosccn100S1.a1, pvcaosccn100M1.a1,
magaosccn100M1.a1
Description:
The variable CCN-temp unstable is incorrectly defined as a hexadecimal value. The values 
are 0 or 1 only. The description is also reversed. A value of 1 indicates the instrument 
is operating stably.
Suggestions: 
Reverse the values. Change 0 to 1, and 1 to 0
Measurements:tmpaosccn100S1.a1:
  • CCN_temp_unstable_flag
more
pvcaosccn100M1.a1:
  • CCN_temp_unstable_flag
more
magaosccn100M1.a1:
  • CCN_temp_unstable_flag
more
pvcaosccn100S1.a1:
  • CCN_temp_unstable_flag
more
enaaosccn100C1.a1:
  • CCN_temp_unstable_flag
more

Close this window