Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
01/09/2015 | 0000 | 02/16/2015 | 0000 | Suspect |
Subject: | ACX/OZONE/M1 - Data noisy while ship is under way |
DataStreams: | acxaoso3M1.a1 |
Description: | Data is noisy at high frequency while ship is under way. Regular daily calibrations show the accuracy is correct, only the noise is severe (+/- 5-10 ppbv). |
Suggestions: | Averaging data for 20 to 30 seconds GREATLY reduces noise. |
Measurements: | acxaoso3M1.a1: more
|