Subject:
ACX/HTDMA/M1 - Failure to fill butanol into HTDMA
DataStreams:acxaoshtdmaM1.a1, acxaoshtdmaM1.00
Description:
CPC counts went to zero and remained at zero until the end of the stated time range. The 
CPC was on and detecting only shot noise from the optical detector. Indications suggest 
that butanol was not being filled into the HT-DMA
Suggestions: 
Do not use data