Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
04/01/2015 | 1513 | 04/07/2015 | 1713 | Suspect |
Subject: | SGP/MFRSR/E34 - Head Temp - Loose cable |
DataStreams: | sgpmfrsrlangleyE34.c1, sgpmfrsrE34.b1, sgpmfrsrE34.a0, sgpmfrsrlangplotE34.c1 |
Description: | A cable was loose causing the head temperature to not be held at 40 degC. The instrument is calibrated and designed to operate at a stable temperature of 40 degC. Coloring this DQR as yellow as the data might be suitable for some users. Contact the mentor for guidance. |
Suggestions: | |
Measurements: | sgpmfrsrE34.b1: more
sgpmfrsrE34.a0: more sgpmfrsrlangplotE34.c1: more sgpmfrsrlangleyE34.c1: more |