Start Date | Start Time | End Date | End Time | Data Quality Metric |
---|---|---|---|---|
02/19/2015 | 0925 | 02/19/2015 | 1105 | Incorrect |
02/16/2015 | 1052 | 02/17/2015 | 0524 | Incorrect |
02/17/2015 | 0557 | 02/17/2015 | 0732 | Incorrect |
04/27/2015 | 0942 | 04/27/2015 | 1523 | Incorrect |
05/03/2015 | 0522 | 05/03/2015 | 0811 | Incorrect |
05/03/2015 | 0227 | 05/03/2015 | 0522 | Incorrect |
02/20/2015 | 1914 | 02/22/2015 | 2131 | Incorrect |
02/19/2015 | 1541 | 02/20/2015 | 0148 | Incorrect |
05/08/2015 | 0942 | 05/08/2015 | 1340 | Incorrect |
02/17/2015 | 0858 | 02/17/2015 | 1135 | Incorrect |
02/18/2015 | 0954 | 02/18/2015 | 1144 | Incorrect |
02/19/2015 | 0808 | 02/19/2015 | 0914 | Incorrect |
04/26/2015 | 0907 | 04/26/2015 | 1352 | Incorrect |
02/20/2015 | 1255 | 02/20/2015 | 1829 | Incorrect |
Subject: | ENA/AOSMET/C1 - Instrument |
DataStreams: | enaaosmetC1.a1 |
Description: | Instrument experienced periodic dropouts. These were apparent on all channels. The RH sensor went off scale. During these periods there are no valid points for Air Temperature, RH or BP. RH should be considered uncalibrated until this date. Changing the sensor at the end of this period fixed the problem. |
Suggestions: | |
Measurements: | enaaosmetC1.a1: more
|